Could standardizing "commercial off-the-shelf" (COTS) monitors to the DICOM part 14: GSDF improve the presentation of dental images? A visual grading characteristics analysis
Date Issued
2013-10-01
Date Available
2025-03-26T16:10:13Z
Abstract
Objectives: To investigate whether standardizing commercial off-the-shelf (COTS) display devices to the digital imaging and communications in medicine part 14: greyscale standard display function (DICOM part 14: GSDF) would affect the presentation of dental images. Methods: Two COTS display devices from the radiology department of a dental teaching hospital and a laptop computer monitor for reference were calibrated to conform to DICOM part 14: GSDF. Four dental surgeons and two final-year students undertook a relative visual grading analysis of the two devices before and after calibration, under control of the viewing environment. Results: Calibrating COTS display devices to conform to the DICOM part 14: GSDF and viewing under reduced ambient light result in a consistent, perceived visual sensation for the presented radiological image. The area under the visual grading characteristics curve (AUCVGC) before calibration is 0.62 CI (0.56, 0.68) and AUCVGC after calibration is 0.51 CI (0.45, 0.57). Conclusions: Standardizing COTS display devices to the DICOM part 14: GSDF can improve image presentation.
Type of Material
Journal Article
Publisher
British Institute of Radiology
Journal
Dentomaxillofacial Radiology
Volume
42
Issue
9
Start Page
1
End Page
6
Copyright (Published Version)
2013 The Authors
Subjects
Language
English
Status of Item
Peer reviewed
ISSN
0250-832X
This item is made available under a Creative Commons License
File(s)
Loading...
Name
Could standardizing commercial off-the-shelf (COTS) monitors to the DICOM part 14 GSDF improve the presentation of dental im.pdf
Size
316.51 KB
Format
Adobe PDF
Checksum (MD5)
23408c4749416707dd06d3b396413a00
Owning collection